Personal Noise Dosimeter SV 104A

The SV 104A is a personal noise dosimeter for assessing occupational sound exposure. It is built to IEC 61252 and ANSI S1.25-1991 (R2020), the standards for personal sound exposure meters, whose acceptance limits correspond to class 2 of ANSI/ASA S1.4 / IEC 61672. Its three profiles work as three independent dosimeters on the same worker, each with its own exchange rate (2, 3, 4, 5 or 6 dB), criterion level and threshold. For example, one measurement run can use any three standards at once — OSHA, ACGIH and NIOSH. The time-history logger records TWA, DOSE, LEX,8h and Leq/Peak/Max/Min. The inbuilt Bluetooth® interface adds two-way remote communication with a smartphone app while the dosimeter is worn, so a run can be started, checked and stopped without approaching the worker. It opens the way to remote noise telemetry.

bestseller-icon
Bestseller
mems-icon
MEMS
oled-icon
OLED

SV 104A Price, Kits and Options

The SV 104A ships ready to measure and costs from EUR 1,500, depending on local taxes, duties and exchange rates.

Price

The instrument alone costs from EUR 1,500 to 2,400, or about USD 1,635 to 2,616. Kits run from EUR 2,000 to 3,400, about USD 2,180 to 3,706. Ask for a quote through the enquiry form on this page or through an authorised SVANTEK distributor.

Kits

The personal noise dosimeter on its own comes with the microphone, windscreen, USB cable, battery and transport box, but no charger. K1 adds a charger, an SV 34B class 2 calibrator, a case and the manual; K5 adds a hub for five units; D5 a docking station for five.

Options

Octave analysis and audio recording are licensed options and can be activated at any time. The triaxial vibration sensor is not an option: it runs in every configuration, with no licence. For explosive atmospheres the SV 104IS is a separate version of the instrument.

SV104A The instrument is built to IEC 61252 and ANSI S1.25-1991 (R2020) SVANTEK Sound and Vibration

Hardware and Software Features

MEMS Microphone

Patented, Shock-Resistant, and Life-Time Warranty

The SVANTEK MEMS microphone is patented and resistant to mechanical shock, as a dosimeter on a worker's shoulder has to be. Its low power draw compared with a piezoelectric sensor is one reason the instrument runs for over 48 hours. It carries a lifetime warranty.
SV104A MEMS Microphone SVANTEK

Frequency Analysis and Audio Recording

Enhanced Noise Source Identification

Optional octave analysis shows the spectrum of a disturbance and threshold-triggered audio recording identifies what caused it, such as a radio or speech unrelated to the job. A spectrum also tells you which hearing protector will attenuate that noise, and a voice note lets the person running the study record what was happening. All of it goes into a single time-history data file, so those periods can be judged for relevance and excluded from the dose.
SV104A Frequency Analysis and Audio Recording SVANTEK

Bluetooth App

Remote Control and Real-Time Monitoring

The Bluetooth® interface allows the measurement to be controlled and read remotely. From the ASSISTANT PRO app on a phone or tablet you start and stop a run and check results without touching the instrument. Touching a running dosimeter is one way to invalidate a measurement. The app raises an alarm when a set level is exceeded.
SV104A Bluetooth App SVANTEK

Auto-Calibration

Simplifying Accurate Measurements

The SV 104A features an auto-calibration function that runs the before-and-after field calibration required by ISO 9612 and stores the calibration data with the measurement. It recognises the calibrator signal on its own, so nobody has to open a menu at the start and the end of a shift. Those are the two moments when the check gets skipped.
SV104A Auto-Calibration SVANTEK

Inbuilt Vibration Sensor

Ensuring Accurate Noise Measurement

Worn on the shoulder for continuous, unattended measurement, the dosimeter detects the false contributions described in ISO 9612. The internal triaxial sensor logs mechanical shock and movement and shows when the microphone was knocked or the instrument taken off. Three axes let it tell a knock from the normal movement of a worker walking between machines, and the sensor runs in every configuration, with no licence to activate.
SV104A Inbuilt Vibration Sensor SVANTEK

SV 104A Datasheet Catalogue

SV 104A Personal Noise Dosimeter

The MEMS microphone carries a lifetime warranty and holds its stability and accuracy in harsh industrial conditions, because it is built to survive shock and drops. The linear operating range to IEC 61672 starts at 53 dBA, and the total range is 43 dBA RMS to 141 dBA Peak.

Noise Dosimeter Practical Applications

Occupational Noise

Noise Measurements for Regulatory Compliance

The SV 104A is used for occupational noise measurement wherever the work is done: under OSHA 29 CFR 1910.95 in the United States, where MSHA, NIOSH and ACGIH add further criteria, and under ISO 9612 against Directive 2003/10/EC in Europe. Because the three profiles are independent, one study can cover a legal limit and a stricter recommendation at the same time, without measuring twice. In construction, manufacturing, mining and oil and gas, workers move between positions and levels vary through the shift, so one dosimeter run replaces a series of spot measurements

Read more
SV104A Occupational Noise SVANTEK

Videos

New generation of noise dosimeters

The story of SV104 noise dosimeter, the first dosimeter with a MEMS microphone

Noise Dosimeter SV 104A

Discover unique features of the SV 104A noise dosimeter

Noise dosimeters application

How loud a Speedway match is? See how the dosimeter is used in practice

Configuration for noise measurement

Use of the Supervisor Software with noise dosimeters

Check Availability & Price














    Products in the enquiry (1)

    I hereby consent to the processing of my personal data, i.e. my full name and e-mail address, by SVANTEK SP. Z O.O. with its registered office in Warsaw at ul. Strzygłowska 81 for the purpose of receiving marketing information on the products and services offered by SVANTEK SP. Z O.O. via electronic means of communication, in particular via e-mail, in accordance with the provisions of Article 10 sec. 1 and 2 of the Act on providing services by electronic means.

    I hereby consent to the processing of my personal data, i.e. my full name and phone number, by SVANTEK SP. Z O.O. with its registered office in Warsaw at ul. Strzygłowska 81 for the purpose of marketing activities with the use of telecommunications terminal equipment and automatic calling machines within the meaning of the Telecommunications Act.

    I hereby consent to receiving from SVANTEK SP. Z O.O. with its registered office in Warsaw at ul. Strzygłowska 81, via electronic means to the e-mail address I have provided, the newsletter and marketing information on the products and services offered by SVANTEK SP. Z O.O., within the meaning of the Act on providing services by electronic means.


    I declare that I have been informed that my data may be transferred to entities that process personal data on behalf of the Administrator, in particular to distributors - such entities process data on the basis of an agreement with the Administrator and exclusively in accordance with its instructions. In such cases, the Administrator requires third parties to maintain the confidentiality and security of information and verifies that they provide appropriate measures to protect personal data.

    Some of the entities processing personal data on behalf of the Administrator are established outside the EEA. In connection with the transfer of your data outside the EEA, the Administrator verifies that these entities provide guarantees of a high level of personal data protection. These guarantees stem in particular from the obligation to apply the standard contractual clauses adopted by the Commission (EU). You have the right to request a copy of the standard contractual clauses by sending a request to the Controller.

    I declare that I have been informed on my right to withdraw my consent to the processing of my personal data at any time, to access the provided personal data, to rectify, erase, restrict processing and object to the processing of my data, as well as the right to lodge a complaint with the President of the of the Personal Data Protection Office in the event of an infringement of the provisions of GDPR.

    SV 104A Specifications

    SV 104A – Personal Noise Dosimeter

    Applications

    Sound measurements
    Occupational noise

    Hardware Features

    Number of channels
    1
    Intrisic safety
    Self-vibration detection
    Auto-calibration
    Remote system check
    Keyboard lock
    Remote communication
    Bluetooth
    Communication Interfaces
    USB 2.0, Bluetooth® 4.0, electrical contacts
    LED indicator
    Display
    Keyboard
    3 push buttons
    Memory
    8 GB
    Battery type
    Li-Ion rechargeable cell
    Operating time
    operation time > 48 hours
    Operating temperature range
    from -10 o C to 50 o C
    Humidity range
    up to 90 % RH
    Weight
    121 grams
    Size
    88 x 49.5 x 19.2 mm
    IP rate
    IP 65
    Docking station required

    Sound measurements features

    Standards
    IEC 61252 Ed 1.2 (R2017); ANSI S1.25-1991 (R2020) Class 2 IEC 61672-1 (2013)
    Microphone type
    ST 104CIS MEMS microphone, 1/2” housing, patented
    Preamplifier type
    Integrated
    Linear operating range to IEC 61672
    53 dBA RMS ÷ 141 dBA Peak
    Total measurement range
    43 dBA RMS ÷ 141 dBA Peak
    Frequency range
    20 Hz ÷ 10 kHz
    Weighting filters
    A, C and Z
    Time constants
    Slow, Fast, Impulse
    Exchange rates
    2, 3, 4, 5, 6
    Criterion levels
    Selectable
    Profiles per channel
    3 with independent settings of filters (x) and time constants (y)
    Min logging step
    1s logger step
    Statistics
    LN (LEQ STATISTICS)
    Statistics in octaves or third octaves
    Audio recording
    option
    Voice coments
    1/1 octave
    option
    1/3 octave
    option

    Downloads