| VIBRATION LEVEL METER / ANALYSER |
|
Standards
|
ISO 8041:2005
|
| Meter mode |
RMS, VDV, MTVV or MAX, Peak, Peak-Peak
Simultaneous measurement in up to four channels with independent set of filters and detector time constants |
| Analyser (option) |
1/1 octave 1 real time analysis, 15 filters with centre frequencies from 1 Hz to 16 kHz (Type 1, IEC 61260)
1/3 octave 1 real time analysis, 45 filters with centre frequencies from 0.8 Hz to 20 kHz (Type 1, IEC 61260)
FFT 1 real time analysis up to 1600 lines with Hanning, Kaiser-Bessel or Flat Top window
FFT 1 cross spectra measurements
RPM 1 rotation speed measurements parallel to the vibration measurement (1 ÷ 99999)
and more... |
| Filters |
Wd , Wk , Wc , Wj , Wm , Wb , Wg (ISO 2631), Wh (ISO 5439),
HP1, HP3, HP10, Vel1, Vel3, Vel10, VelMF, Dil1, Dil3, Dil10, KB |
| RMS & RMQ Detectors |
Digital True RMS & RMQ detectors with Peak detection, resolution 0.1 dB
Time Constants: from 100 ms to 10 s |
| Accelerometer (option) |
SV 39A/L seat accelerometer (100 mV/g sensitivity) for Whole-Body measurements
SV 50 set for Hand-Arm measurements (Dytran 3023M2 accelerometer 10 mV/g sensitivity)
Other IEPE accelerometers |
|
Measurement Range
|
accelerometer dependent, 0.003 ms-2 RMS ÷ 500 ms-2 PEAK (with 100 mV/g sensitivity)
|
| Frequency Range |
0.5 Hz ÷ 20 kHz; accelerometer dependent
with Dytran 3023M2: 2 Hz ÷ 10 kHz
|
| SOUND LEVEL METER / ANALYSER |
|
Standards
|
Type 1: IEC 61672-1:2002
|
| Meter mode |
SPL, Leq , SEL, Lden , Ltm3 , Ltm5 , Statistics - L n (L 1 -L 99 ), LMax , LMin , LPeak
Simultaneous measurement in three profiles per channel (up to four channels) with independent set of filters and detector time constant |
| Analyser (option) |
1/1 octave 1 real time analysis, 15 filters with centre frequencies from 1 Hz to 16 kHz (Type 1, IEC 61260)
1/3 octave 1 real time analysis, 45 filters with centre frequencies from 0.8 Hz to 20 kHz (Type 1, IEC 61260)
Reverberation Time analysis in 1/3 octave bands (RT 60)
FFT 1 real time analysis up to 1600 lines with Hanning, Kaiser-Bessel or Flat Top window
FFT 1 cross spectra measurements
Sound Intensity Measurements
and more. |
| Weighting Filters |
standard: A, C, Z and G |
| RMS Detector |
Digital True RMS detector with Peak detection, resolution 0.1 dB
Time Constants: Slow, Fast, Impulse |
| Microphone (option) |
SV 22, Type 1, 50 mV/Pa, prepolarised 1/2" condenser microphone with SV 12L preamplifier
SV 25, Type 2, dosimeter microphone, ceramic, 1/2" casing with integrated preamplifier |
| Measurement Range |
Total Dynamic Range: 17 dBA RMS ÷ 140 dBA Peak (with 50 mV/Pa microphone sensitivity)
Linearity Range: 24 dBA RMS ÷ 140 dBA Peak (with 50 mV/Pa microphone sensitivity) |
| Frequency Range |
0.5 Hz ÷ 20 kHz; microphone dependent,
with SV 22 microphone: 10 Hz ÷ 20 kHz |
| BASIC DATA |
|
Input
|
IEPE type (channels 1, 2, 3 - LEMO 4 pin & channel 4 - TNC connector)
|
| Dynamic Range |
100 dB, 4 x 20 bits A/D converters |
| Frequency Range |
0.5 Hz ÷ 20 kHz, sampling rate 48 kHz |
| Data Logger 1 |
Time History logging to internal memory or external USB Memory Stick
Time domain signal recording on USB Memory Stick (option) |
| Display |
LCD 128 x 64 pixels plus icons with backlighting |
| Memory |
32 MB non-volatile flash type, external USB Memory Stick (not included) |
| Interfaces |
USB 1.1 Client, USB 1.1 Host, RS 232 (option: SV 55 requiered), IrDA (option)
Extended I/O - AC output (1 V Peak) or Digital Input/Output (Trigger - Pulse) |
| Power Supply |
| Four AA batteries (alkaline) |
operation time > 10 h
(6.0 V / 1.6 Ah)2 |
| Four NiMH rechargeable batteries (not included) |
operation time > 14 h
(4.8 V / 2.6 Ah)2 |
| SA 17A external battery pack (option) |
operation time > 24 h 2 |
| External power supply (SA 15 not included) |
6 V DC ÷ 15 V DC (1.5 W) |
| USB interface |
500 mA HUB |
|
| Environmental Conditions |
| Temperature |
from -10oC to 50oC |
| Humidity |
up to 90 % RH, non-condensed |
|
| Dimensions |
140 x 82 x 42 mm |
| Weight |
510 grams with batteries |
1each function parallel to the meter mode